電磁界の生体影響を考慮した表皮平行型経皮トランス- 受電コイルを直並列に配置した場合-

Takuya Fujitsuka, Kenji Shiba

研究成果: Article査読

抄録

In this study, we created a skin-parallel type transcutaneous transmission system for a ventricular assist device using plate-shaped solenoid as the transmitting coil. We compared the biological effects and radiated magnetic fields of six models in which two receiving coils were placed in series and in parallel. The results showed that the power transmission efficiency was around 90% for all models, and there was no major difference between the two layouts. It was also found that by placing two receiving coils in parallel and separating them by 10 mm (Type D), was possible to suppress the electric field strength within the body Ein by more than 10 V/m. However, the limit value was exceeded by placing two receiving coils in series and separating them by 0.5 mm. In addition, the local SAR for Type D was 1.2595 W/kg, which was below the limit value.

寄稿の翻訳タイトルSkin-Parallel Type Transcutaneous Energy Transmission Transformer Considering the Biological Effects of Electromagnetic Fields -When Receiving Coils are Arranged in Series and Parallel-
本文言語Japanese
ページ(範囲)399-407
ページ数9
ジャーナルIEEJ Transactions on Electronics, Information and Systems
145
3
DOI
出版ステータスPublished - 1 3月 2025

Keywords

  • energy transmission efficiency
  • ferrite core
  • SAR
  • stimulant effect
  • TETS